12.
EMC
Guidance and manufacture's declaration – electromagnetic emission
The RBP-100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor is intended
for use in the electromagnetic environment specified below. The customer of the user of the RBP-
100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor should assure that it
is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacture's declaration – electromagnetic immunity
The RBP-100 Blood Pressure Monitor and the RBP-100 USB Blood Pressure Monitor is intended for use in
the electromagnetic environment specified below. The customer or the user of RBP-100 Blood Pressure
Monitor and the RBP-100 USB Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Electrostatic
Con:±8 kV
discharge (ESD)
Air:±2,4,8,15 kV
IEC 61000-4-2
Electrical fast
Pass
transient/burst
IEC 61000-4-4
Surge
Pass
IEC 61000-4-5
Voltage dips, short
Pass
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
30A/m
(50Hz/60Hz)
magnetic field IEC
61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 80601-2-30
EN 1060-1 /-3
Compliance
Electromagnetic environment – guidance
The RBP-100 Blood Pressure Monitor and the
RBP-100 USB Blood Pressure Monitor RF
energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
The RBP-100 Blood Pressure Monitor and the
Class B
RBP-100 USB Blood Pressure Monitor is
suitable for use in all establishments, other
than domestic and those directly connected to
Pass
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Pass
Compliance level
Con:±8 kV
Air:±2,4,8,15 kV
Pass
Pass
Pass
30A/m
Seite 19 von 23
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the RBP-100 Blood Pressure
Monitor and the RBP-100 USB
Blood Pressure Monitor requires
continued operation during power
mains interruptions, it is
recommended that the RBP-100
Blood Pressure Monitor and the
RBP-100 USB Blood Pressure
Monitor be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
RBP-100