12.  
EMC  
Guidance  and  manufacture's  declaration  –  electromagnetic  emission  
The  RBP-­100  Blood  Pressure  Monitor  and  the  RBP-­100  USB  Blood  Pressure  Monitor  is  intended  
for  use  in  the  electromagnetic  environment  specified  below.  The  customer  of  the  user  of  the  RBP-­
100  Blood  Pressure  Monitor  and  the  RBP-­100  USB  Blood  Pressure  Monitor  should  assure  that  it  
is  used  in  such  an  environment.  
Emission  test  
RF  emissions  
CISPR  11  
RF  emission  
CISPR  11  
Harmonic  emissions  
IEC  61000-­3-­2  
Voltage  fluctuations/  
flicker  emissions  
 
IEC  61000-­3-­3  
 
 
 
 
Guidance  and  manufacture's  declaration  –  electromagnetic  immunity  
The  RBP-­100  Blood  Pressure  Monitor  and  the  RBP-­100  USB  Blood  Pressure  Monitor  is  intended  for  use  in  
the  electromagnetic  environment  specified  below.  The  customer  or  the  user  of  RBP-­100  Blood  Pressure  
Monitor  and  the  RBP-­100  USB  Blood  Pressure  Monitor  should  assure  that  it  is  used  in  such  an  
environment.  
Immunity  test  
IEC  60601  test  level  
Electrostatic  
Con:±8  kV  
discharge  (ESD)  
Air:±2,4,8,15  kV  
IEC  61000-­4-­2  
 
Electrical  fast  
Pass  
transient/burst  
IEC  61000-­4-­4  
Surge  
Pass
IEC  61000-­4-­5  
Voltage  dips,  short  
Pass
interruptions  and  
voltage  variations  on  
power  supply  input  
lines  
IEC  61000-­4-­11  
Power  frequency  
30A/m  
(50Hz/60Hz)  
magnetic  field  IEC  
61000-­4-­8  
NOTE      U
 is  the  a.c.  mains  voltage  prior  to  application  of  the  test  level.  
T
 
 
 
 
 
 
 
 
 
 
 
IEC  80601-­2-­30  
EN  1060-­1  /-­3  
Compliance  
Electromagnetic  environment  –  guidance  
The  RBP-­100  Blood  Pressure  Monitor  and  the  
RBP-­100  USB  Blood  Pressure  Monitor  RF  
energy  only  for  its  internal  function.  Therefore,  
Group  1  
its  RF  emissions  are  very  low  and  are  not  
likely  to  cause  any  interference  in  nearby  
electronic  equipment.  
The  RBP-­100  Blood  Pressure  Monitor  and  the  
Class  B  
RBP-­100  USB  Blood  Pressure  Monitor  is  
suitable  for  use  in  all  establishments,  other  
than  domestic  and  those  directly  connected  to  
Pass  
the  public  low-­voltage  power  supply  network  
that  supplies  buildings  used  for  domestic  
purposes.  
Pass  
 
 
Compliance  level  
Con:±8  kV  
Air:±2,4,8,15  kV  
 
Pass  
Pass
Pass
30A/m  
Seite 19 von 23
Electromagnetic  environment  -­  
guidance  
Floors  should  be  wood,  concrete  
or  ceramic  tile.  If  floor  are  
covered  with  synthetic  material,  
the  relative  humidity  should  be  at  
least  30%.  
Mains  power  quality  should  be  
that  of  a  typical  commercial  or  
hospital  environment.  
Mains  power  quality  should  be  
that  of  a  typical  commercial  or  
hospital  environment.  
Mains  power  quality  should  be  
that  of  a  typical  commercial  or  
hospital  environment.  If  the  user  
of  the  RBP-­100  Blood  Pressure  
Monitor  and  the  RBP-­100  USB  
Blood  Pressure  Monitor  requires  
continued  operation  during  power  
mains  interruptions,  it  is  
recommended  that  the  RBP-­100  
Blood  Pressure  Monitor  and  the  
RBP-­100  USB  Blood  Pressure  
Monitor  be  powered  from  an  
uninterruptible  power  supply  or  a  
battery.  
 
Power  frequency  magnetic  fields  
should  be  at  levels  characteristic  
of  a  typical  location  in  a  typical  
commercial  or  hospital  
environment.  
RBP-100