9.4.1.8
Please wait until this screen appears.
9.4.1.9
Example for the correct value reading.
9.4.1.20 Pump via the Hand Inflation Bulb (4) up to 300mmHg.
Compare the pressure displayed on the screens of the device and the calibrated
reference manometer.
9.4.1.21 Then carefully release the air via the Hand Inflation Bulb (4) and stop
at the value of 250mmHg. Compare the pressure displayed on the screen of the
device and the calibrated reference manometer.
Note the displayed value on the RBP-100.
Note the displayed value at the calibrated reference manometer.
9.4.1.22 Please repeat these steps at the values of 200-150-100-50-0mmHg.
Note the displayed value on the RBP-100.
Note the displayed value at the calibrated reference manometer.
9.4.1.23 Please check the readings on the RBP-100 against the value according to the
manometer values shown in Table 1 below.