8.

Troubleshooting list

8.1

Bonum battery operated clipping machine

Malfunction
The battery-operated clipping machine does not
start
The battery-operated clipping machine pauses dur-
ing operation
8.2

Charger and Li-ion battery

Malfunction
Charger not working
Li-ion battery cannot be loaded
Cause
Li-ion battery not loaded
Li-ion battery defect
Li-ion battery not inserted
Li-ion battery deactivated through protective
circuit
Li-ion battery is empty
Detection
LED, operation indicator d does
not illuminate
Li-ion battery not loaded
Li-Ion battery inserted, LED,
operation indicator d does not
light up
Li-Ion battery inserted, LED,
warning indicator e lights up
continuously
Li-Ion battery inserted, LED,
warning indicator e blinking
Remedy
Load Li-ion battery
Insert new Li-ion battery
Insert the Li-ion battery
Switch the battery-operated clipping machine off
and on again
Insert new Li-ion battery
Cause
Mains power cord not plugged in
Mains power cord defective
Li-ion battery defect
Charger contacts dirty
Charger contacts damaged
Li-ion battery defect
Charger defect
Too high a temperature (>50°) is
measured in the Li-Ion battery dur-
ing the charging process
Battery charging error, foreign
object on contact system or Li-Ion
battery defective
Remedy
Plug in the mains connection line at
the mains power socket.
Send charger to manufacturer for
repair; see Technical Service
Send Li-ion battery to manufacturer
for repair, see Technical Service
Clean the contacts in the charging bay,
see Reprocessing
Send charger to manufacturer for
repair, see Technical Service
Send Li-ion battery to manufacturer
for repair, see Technical Service
Send charger to manufacturer for
repair, see Technical Service
Remove the Li-ion battery from the
charging bay, leave to cool and repeat
the charging process
If the problem occurs again, have Li-
Ion battery inspected by manufacturer,
see Technical Service
Remove the Li-Ion battery from the
charging bay, clean contact system
and repeat charging process
If the problem occurs again, have Li-
Ion battery inspected by manufacturer,
see Technical Service
en
9

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis