HBM TB1A Montageanleitung Seite 16

Referenz-drehmomentmessscheibe
Vorschau ausblenden Andere Handbücher für TB1A:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

16
Installing as a transfer transducer
Transfer transducers must be as insensitive as possible to all installation
influences. This can be achieved in the design, for example, by specially
developed adaptation flanges.
For optimum transfer of sensitivity, observe the following points in addition to
the recommendations for reference transducers already mentioned:
• Introduce the torque into the torque measuring disc from the inside (D
the outside (D
A
• The width of the adaptation flange (B) on the reaction side should be 1.5 −
2 times the diameter of the flange screw.
TB1A measuring side
Fig. 4.5: Adaptation flange transfer transducer
HBM
D
I
), at a ratio of
D
A
Adaptation flanges
v 0.6 .
TB1A reaction side
B
TB1A
) to
I
D
A
D
I
A0162 7.0 en/de/fr