• Contamination of the inside of the pipette caused by
extensive aspiration of chemically aggre ssive liquids or
because liquid got inside the pipette.
Remove the tip ejector. Unscrew the nut, remove
the shaft, piston assembly, seal and O-ring. Rinse
the removed parts with distilled water and dry thor-
oughly. Lightly lubricate the inside surfaces of the
seal and the O-ring with the lubricant. Reassemble
the pipette in the reverse order.
Note: The shaft, shaft nut, ejector, piston assem bly,
seal and O-ring may be autoclaved at a temperature
of 121°C for 20 minutes at pressure 1 bar.
The shaft of the 5000 and 10000 models should be
autoclaved without the filter.
The handle and the outside of the pipette may be
cleaned using a cloth damped in isopropanol.
Warning: Before using cleaning agents other than those
recommended by the manufacturer, check the compati-
bility charts and consider chemical resistance of the fol-
lowing plastics: PP, PC, POM, PA, PPS, PVDF used to
produce the pipette parts.
If the pipette malfunction persists after carring out the
above steps, send the pipette to your HTL service repre-
sentative.
Before returning the pipette, make sure that it is free from
any chemical, radioactive or microbiological contamina-
tion that might pose any danger during transportation and
repairing. Clean the pipette as thoroughly as possible.

10 - PIPETTE KIT

The pipettes are delivered in the kits including:
• Pipette
• Short instruction manual
• Calibration key
• Ejector cap (for pipette models OP10000)
• Ejector spacers (for pipette models OP2-OP1000)
• CD
13
Note:
The following parts are not part of the standard equip-
ment of the pipette and are available upon request:
• Ejector cap (for pipette models OP2, OP10, OP5000)
• Pipette stand
• Lubricant
• Filters (for pipette models OP5000, OP10000)
• Color identification rings (for pipette models OP2-
OP1000)

11 - SPARE PARTS AND ACCESSORIES

All the spare parts and accessories indicated in Fig. 3, 4,
5 and 6:
A: Pushbutton
B: Adjustment knob
C: Shaft
D: Ejector
F: Shaft nut
G: Piston assembly
I: O-ring
J: Seal
K: Calibration key
L: Filter
M: Ejector cap
N: Ejector pushbutton
P: Identification ring
S: Pipette stand
The stand assembly diagram is shown in Fig. 5.
Spare parts and accessories can be ordered from your
HTL representative (type of pipette and name of the part
for this pipette should be specified).
Warning: The replacement of the piston requires
conducting of calibration procedure according to
section 8.
All rights reserved. Product described in this manual is subject to availability
and technical modification. Errors excepted.
PZ HTL S.A. reserves the right to improve, enhance or otherwise modify its
products without prior notification.
© 2015 PZ HTL S.A.