If the average aspirated volume exceeds the per missible
value, the following should be done:
• Remove the pipetting pushbutton, (Fig. 4A),
• Holding the volume setting knob to protect it against
rotation, insert the calibration key into the cuts of the
calibration screw, (Fig. 4B),
• Turn the key clockwise to reduce the aspirated volume,
or counterclockwise to increase the volume. One full
turn of the calibration key changes the pipette aspira-
tion volume by the amount given in the table, (Fig. 4C),
• Take out the key and fix the pipetting pushbutton,
(Fig. 4D).
Determine the average aspirated volume. The average
volume should be within the permissible range given in
the table. If the volume exceeds the values stated, the
recalibration procedure should be repeated.
More information on calibration procedure can be found
on www.htl.com.pl

6 - CLEANING AND STERILIZATION

Cleaning
External surfaces of the pipetting pushbutton, the ejector
button, the handgrip, shaft and the adjustment knob may
be cleaned using a cloth dampened in isopropyl alcohol.
Cones should be kept clean using isopropyl alcohol.
Warning: Before using cleaning agents other than those
recommended by the manufacturer, check the compati-
bility charts and consider chemical resistance of the fol-
lowing plastics: PP, PC, POM, PA, PPS, PVDF used to
produce the pipette parts.
7
Pressure [kPa]
95.0
101.3
105.0
1.0028
1.0029
1.0029
1.0030
1.0031
1.0031
1.0032
1.0033
1.0033
1.0034
1.0035
1.0036
1.0037
1.0038
1.0038
1.0039
1.0040
1.0040
The pipettes are UV resistant which was confirmed by
our tests. The recommended distance from the radiation
source to exposed element should be not less than
50 cm. Prolonged and very intense UV exposure can
cause de-coloration of pipette parts, without affecting
its performance.

7 - PIPETTE KIT

The pipettes are delivered in the kits including:
• Pipette
• Short instruction manual
• Calibration key
• CD

8 - SPARE PARTS AND ACCESSORIES

The spare parts for multichannel in, (Fig. 1, 4):
A: Pipetting pushbutton
B: Adjustment knob
C: Ejector button
D: Piston's assembly
E: Shaft
F: Ejector
J: Calibration key
Warning: The replacement of the plunger requires
conducting of calibration procedure according to
section 5.
Before returning a pipette to us for service, please
ensure that the pipette is completely free of any chemi-
cal, biological or radio active contamination or with the
information on the kind of liquids that have been mea-
sured.
All rights reserved. Product described in this manual is subject to availability
and technical modification. Errors excepted.
PZ HTL S.A. reserves the right to improve, enhance or otherwise modify its
products without prior notification.
© 2015 PZ HTL S.A.
ENGLISH
8

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis