3.
Maintain a distance of no more than 10 feet (3m)
between the base unit and foot control during the
synchronization process.
4.
Remove any inserts from the handpiece and adjust
the Power Level Control out of Rinse Mode. Turn the
Main Power switch to the ON (I) position and wait for
the Diagnostic Display graphics to light (refer to
Section 8.2).
5.
While all graphics are lit, press the Purge button,
located on the Diagnostic Display.
The graphics will begin to blink in a sequential
pattern, representing the synchronization mode.
This mode will last 5 to 6 seconds.
6.
During this mode, press the red button located in the
battery compartment of the foot control. This will
complete the synchronization process.
7.
Synchronization is successful when all graphics blink
at the same time.
8.
To verify proper communication, press the foot
control to the Boost position (foot control fully
depressed – 2
position) and ensure the Boost
nd
graphic on base unit lights.
9.
Replace battery cover and the screw.
10.
In the event communication cannot be established,
temporarily use the supplied Auxillary Foot Control
Cable to connect the Foot Control directly to the unit.
9