5.8.2 Linearization of non verifiable models (low resolution) and verifiable
models
Balance display
In case of verifiable models press the adjustment switch on the lower side of the
balance!
18
Operation
 Start balance by pressing
carries out a self test, press
appears
 Press
repeatedly until "tECH" appears
 Press
, „Pin" appears
 Press
,
appears
 Press
anew, „Pin" appears again
 Press successively
appears; after the apparition of the stability display
and the signal sound, „LoAd 1" will be displayed
 Put on weight 1 (see table chap.6.8.1)
 Wait for the stability display and the signal sound
 „LoAd 2" appears
 Remove weight 1 and put on the weight 2
 After the stability display and the signal sound „LoAd
3" appears; remove weight 2 and put on weight 3
 After the stability display and the signal sound the
balance carries out a selftest. Remove the weight
during selftest. The balance changes into weighing
mode. Now the linearization is concluded
successfully.
O N