Reagent cartridges are for single-use only. Do NOT refill. Any refilling
or reusing of reagent cartridges will void all Analyzer and parts
warranties and nullify any performance claims.
Keep the OPAQUE type reagent cartridge in an upright position
before attaching and detaching the reagent supply line to avoid
exposure to the chemical reagent. Save the plug to later reinsert it
into the nozzle inlet after removing the cartridge from the Analyzer
during a cartridge replacement process.
To protect against accidental exposure to ultra-violet radiation, do
not operate the UV lamp outside of its protective housing.
(On-Line and Portable TOC Analyzers) Water in the iOS System may
be hot. Before inserting a vial into the iOS System after operating in
On-Line mode, slide the door open and wait 30 seconds to allow
sample to completely drain. Inserting a vial before draining can result
in hot water spray projecting upward out of the iOS System.
The iOS System and vial ports contain sharp needles designed to
pierce the septa of sample vials. Do NOT put fingers or inappropriate
materials into the iOS System or vial port.
When servicing parts inside the Analyzer, ensure that power to the
Analyzer is OFF, and keep hands clear of the reagent syringe
assemblies. The syringes are controlled by moving parts that can
pinch skin.
(Laboratory and Portable TOC Analyzers) For continued protection
against fire hazard, replace fuse with same type and rating.
To avoid potentially dangerous shock, do not touch anything inside
the Analyzer while observing the sample pump.
This procedure involves temporarily installing an Ethernet cable from
the back of the instrument into an internal Ethernet port and should
ONLY be performed by a qualified technician.
e
Sievers M9/M9
TOC-Analysatoren
DLM 77000-05 DE Rev. A
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
W
ARNINGS
295 / 392
S
ICHERHEITSWARNHINWEISE
Betriebs- und Wartungsanleitung
GE Analytical Instruments © 2016