1. Select the desired mode (S/cm) by pressing MODE. The display will immedi-
ately show the measured value according to the previous calibration. Should
you want to recalibrate, press CAL.
2. The instrument permits to choose between the standards in memory (1413
µS/cm, 12.88 mS/cm, 111.8 mS/cm + max. 3 user tables). Select the proper
values and press OK. The unused standards should be switched off.
3. After rinsing the electrode several times with the first standard solution,
immerse it in that standard. The solution temperature is not so critical but
should lie between 0°C and 30°C. When no Pt1000 is used, do not forget
to compensate manually first! Select [CALIBRATE], press OK and follow the
in struc tions on the screen until the calibration is finished.
4. Rinse the electrode several times with the sample, immerse it in that solution
and read the display.
5. Rinse the electrode always after use and store it in distilled water (add some
detergent to keep the spongy platinum surface in perfect condition).
When starting a titration, hold MODE pressed and press
range and avoid cross-over errors due to differences in the measuring fre-
quency. Hold MODE pressed and press
MODE
MODE pressed and press
uring mode.
Temperature compensation:
1. Press CAL.
2. Select [REF.] and press OK.
3. Select the temperature to which all future mea sure ments will be refer red to
(25°C or 20°C) and press OK.
Capacitive compensation:
1. Capacitive compensation in creases accuracy in the very low conductivity rang-
es (<10 µS/cm). Select whether or not this compen sa tion should be applied.
Verify if the attached electrode is completely dry and press CAL.
2. Select [CAP.COMP] and press OK.
3. Select [YES] or [NO] and press OK. Follow the in struc tions on the screen.

Resistivity measurement:

1. Select the desired mode (Ω.cm) by pressing MODE. The display will immedi-
ately show the measured value.
Salinity measurement:
1. Select the desired mode (SAL) by pressing MODE. The display will immediately
show the measured value according to the Practical Salinity Scale (PSS-78)
which is based upon standard seawater at 15°C.
TDS measurement:
1. Select the desired mode (mg/l)(TDS) by pressing MODE. The display will imme-
diately show the measured value.
To browse between conductivity, resistivity, salinity or TDS, hold MODE
pressed and press é sequencially.
pressed and press
C6010 • C6020 • C6030
C6010 • C6020 • C6030
C6010 • C6020 • C6030
C6010 • C6020 • C6030
MODE pressed and press
MODE
ê to return again to the normal meas-
sequencially.

Conductivity measurement

Conductivity measurement
ê to lock the actual
4

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

C6020C6030

Inhaltsverzeichnis