digital interface
RS 232 (pH 325 only)
Environmental conditions
ambient temperature
relative humidity
EMC
Interference transmission
Disturbance
Power supply
Operation time
Energy reduction by automatic switch-off after approx. 1 h
Data storage in case of battery exchange
pH calibration values, measuring values and operation mode are stored
Housing
dimension / weight
Instrument safety
safety standard
degree of protection
48
Technical data
4-pin socket with pin contacts for
RS 232-cable AK 325/B
printer cable AK 325/S
serial interface
data output, no separate excitation
baud rate:
data bits:
stop bit:
parity:
hand shake:
max. cable length: 15 m
output format:
operation:
- 10°C ... + 55°C
storage:
- 25°C ... + 65 °C
yearly mean:
< 80 %
30 days/year:
< 100 %
other days:
< 90 %
Technical fundamental standard EN 50081-1
FCC class A
Technical fundamental standard EN 50082-2
Namur recommendation (standard requirements)
4 x 1.5 V non rechargeable batteries, Typ AA
or
4 x 1.2 V NiCd, Typ AA rechargeable batteries
Battery charger:
230 VAC: Friwo FW1199/11.7864,
120 VAC: Friwo FW1199/11.7880,
5000 operating hours with alkali/ manganese
batteries
or
1500 operating hours with NiCd rechargeable
batteries
172 x 80 x 37 mm / approx. 0.3 kg
IEC 1010 / protection class 3
IEC 529 / IP 66
BA12204/05.95/Pap/pH323-325-SET-2
or
4800
8
1
none
RTS/CTS Handshake
ASCII-strings
Friwo Part. No. 1762613
Friwo Part. No. 1794043