LoBat
No message
Indication range
exceeded
No message
Wrong meas.
values
Instrument doesn't
react upon key
operation
to
to
42
Batteries/accumulators nearly
discharged
Electrode:
•
Diaphragm contaminated
•
Membrane contaminated
Meas. solution:
•
ph-value not stable
•
Temperature not stable
Electrode + Meas. solution:
•
Too low conductivity
•
High temperature
Organic liquids
Electrode:
•
Improper electrode
•
Temperature difference
between calibration and
meas. solution too high
•
Interference voltages
•
Improper meas. procedure
Undefined operation
condition of the processor e.g.
after non-admissible ESD-
admission
Timeout of the serial interface Check connected instrument
Trouble Shooting
Change batteries or charge
accumulator
Electrode:
Clean diaphragm
Clean membrane
Meas. solution:
Possibly temper
Possibly temper
Use suitable electrode
Use suitable electrode
Use suitable electrode
Electrode:
Select suitable electrode
Temper buffer or meas.
solutions
Don't connect PC or printer if
meas. solution is grounded
Attend to special procedures
Press key RCL and switch on
the instrument