Übertragen von Messergebnissen mit Modbus
10.2 Modbus-Parameter ändern
● "MbValueTypes" zeigen die auf Modbus darstellbaren Datentypen.
MbValueType
BIT
BIT
W1
W1
W2
W2
W2 ENRON
R1
R1
R2
R2
R2 ENRON
134
Interpretation
1 Bit
1 Bit
16 bit integer
(signed or unsigned)
16 bit integer
(signed or unsigned)
32 bit integer
(signed or unsigned)
32 bit integer
(signed or unsigned)
32 bit integer
(signed or unsigned)
16 bit floating point
(scaled EUHI not IEEE)
16 bit floating point
(scaled EUHI not IEEE)
32 bit IEEE floating point
32 bit IEEE floating point
32 bit IEEE floating point
Zugriffsart
Modbus Protokoll Typ
(Write Access)
FALSE
1 Single Input
TRUE
1 Coil
FALSE
1 Input Register
TRUE
1 Holding Register
FALSE
2 Input Register
TRUE
2 Holding Register
TRUE
1 Holding Register
FALSE
1 Input Register
TRUE
1 Holding Register
FALSE
2 Input Register
TRUE
2 Holding Register
TRUE
1 Holding Register
Bedienhandbuch, 06/2014, A5E01428662-04
Bediensoftware CVControl