SAFETY INSTRUCTIONS
fibre when the appliance is in use.
• Do not place the distal end of the light cable or of the endoscope directly on the patient or
on any other inflammable matter (linen, gauze, operative field, etc.) because it can be very
hot and cause burns.
• A multisocket or extension cable must not be connected to the EM system.
• The light source is intended for professional use in an operating theatre.
• Before you connect the appliance, check whether the mains voltage and frequency indi-
cated on the appliance correspond with the values of the mains power supply.
• This light source can be connected only to the Sopro camera.
• Do not drop the appliance. If the appliance is dropped, do not reconnect it but return it to
your authorized distributor or directly to the after-sales department.
After use, be sure to remove the light cable carefully without touching the distal tip
which can be very hot and cause burns.
It is advisable to have a second light source available in the operating room to use if an
absence or degradation of performance is observed.
This product is equipped with Group 1 LEDs according to the standard IEC 62471. Do not
look directly at the light to avoid any risk of eye damage.
The potential equalization plug located behind the medical device may be used to equalize the
earth potential of the medical device with that of all the devices connected to the mains in the
environment. Use the common earth of the hospital or the building.
Connect the potential equalisation cable to the equipotential C3 plug on the back of the ap-
pliance.
The devices that connect to the inputs/outputs must comply with IEC 60950-1