SAFETY INSTRUCTIONS

Read the user manual.
• The appliance should only be connected to a supply network with a grounding system.
• Observe the conditions of use and storage.
• The appliance must be opened exclusively by a competent technician authorized by the
manufacturer.
• Do not insert any metal objects into the appliance to avoid any risk of electrocution, fire,
short circuit or dangerous emissions.
• Do not expose the appliance to splashes of water or store in a damp place.
• Only use the accessories delivered with the appliance or offered as an option by the manu-
facturer.
• Do not place any heavy objects on the appliance.
• This unit is not sterile.
• If the power cord is damaged, immediately turn the appliance off. It is dangerous to operate
this appliance with a damaged cable.
• To disconnect the cable, pull it out by the plug. Never pull on the cable itself.
• Disconnect the appliance from the mains if you do not plan to use for a few days or more.
• In the area of installation, the presence of high interference fields (strong magnetic fields for
example) should be avoided. These interference fields can result in malfunctions of the light
source.
• Be sure to provide sufficient air circulation to avoid overheating inside the appliance: at least
15 cm around the appliance. Do not cover the side and back air vents and ensure that the
legs of the device are there.
• Use only the disinfectant methods recommended in section 8.
• Before each use, make sure that the appliance has no rough surface, cutting edge, or protu-
berances which could cause problems for safety.
• The surface temperature of the light cable ends can exceed 41°C (after a few minutes of
use). Therefore, avoid holding area in contact with the patient.
• To avoid any risk of electric shock, this appliance must be connected only to a grounded
mains supply socket.
• This appliance may not be modified without authorization of the manufacturer. - If the medi-
cal device is modified, an inspection and a test must be performed to ensure that the medical
device complies with the safety instructions.
• Check the compatibility of your endoscope with the manufacturer before use.
• This appliance must be used on individuals (patients) suitable for undergoing a procedure
of endoscopy;
• Never look at the light output or the end of the light cable.
• Do not insert anything other than a light cable into the housing [C1]. Doing so may damage
the optical system.
• The use of other cables or accessories than those specified may cause an increase in emis-
sions or decreased immunity of the appliance.
• After use of the appliance, when the fibre is removed from its light-guide, the temperature on
the metal coupler of the fibre is very high and can cause burns.
• The light power at the output of the fibre can cause eye injuries. Caution when handling the