T21P Indicator

4.3 User Settings Menu

Enter the Menu Mode (Section 4.1) and press
Press ENTER to accept the displayed setting.
Parameters
F1-0
[
x]
F1-1
[
x]
F1-2
[
x]
F1-3
[
x]
F1-4
[
x]
F1-5
[
x]
F1-6
[
x]
Calle 23B No. 81A - 62 Modelia PBX: (571) 410 28 48 Fax: (571) 263 72 24 Bogota, D.C. - Colombia.
www.tecnipesaje.com
Description
Settings of "x"
Zero Range
0= 2% of max (default)
1= 5% of max
2= 10% of max
3= 20% of max
Zero Tracking
0= 0 division
Range
1= 0.5 division (default)
2= 1 division
3= 2 division
4= 3 division
5= 4 division
6= 5 division
Initial Zero
0= 10% of max
1= 20% of max
2= 50% of max (default)
3= 100% of max
Measuring
0= kg (default)
Unit
1= lb
Filter Level
0= Low
1= Medium (default)
2= High
3= Maximum
Stability
0= Fast
Indication
1= High (default)
2= Medium
3= Slow
Baud Rate*
0= 9600 (default)
1= 4800
2= 2400
3= 1200
tecnipesaje@tecnipesaje.com
SET-F1
to display [
tecnipesaje@outlook.com
EN-11
].