Kalibrierung anzeigen/drucken

Zeigt einen Kalibrierungsbericht an. Dieser Bericht enthält so viele Seiten wie selektierte
LIBERO W.
Menüleiste - Drucken
Ein neues Gerät ohne optionale Kalibrierung wird mit einem Validierungszertifikat von
ELPRO-BUCHS AG ausgeliefert und kann mit "Kalibierung anzeigen" eingesehen werden.
Validation Certificate
Certificate Number:
VC-83000028
Device Type:
LIBERO WSI
Device ID:
83000028
Device Revision:
7.2
ELPRO performs a 100% temperature calibration of all sensors plus a 100% electrical calibration of all electronics.
Additionally the measurement accuracy and functionality is verified and documented with a system calibration of a
representative sample of LIBEROs after final assembly. ELPRO certifies that the LIBERO mentioned above has been
manufactured according to ELPRO's Quality Assurance procedures and has passed the following tests:
Test and method
Temperature calibration of all sensors (100%) at 0 °C
Temperature calibration of all sensors (100%) at 15 °C
Temperature calibration of all sensors (100%) at 25 °C
Electrical calibration of all electronics (100%) at -35 ° C
Electrical calibration of all electronics (100%) at 0 °C
Electrical calibration of all electronics (100%) at 50 °C
Temperature calibration of assembled LIBERO WSI (representative sample) at - 35 °C
Temperature calibration of assembled LIBERO WSI (representative sample) at 0 °C
Temperature calibration of assembled LIBERO WSI (representative sample) at 25 °C
Temperature calibration of assembled LIBERO WSI (representative sample) at 50 °C
Calibration standards are calibrated by an ISO/IEC 17025 accredited laboratory and are traceable to ISO/IEC 17025,
and/or national/international standards (equivalent to NIST, UKAS, DAkkS).
All certificates for the used calibration standards are on file at ELPRO.
This is a non-signed document produced (and reproduceable) by a validated system.
KALIBRIERUNG
Acceptance
Result
criteria
±0.3 °C
passed
±0.3 °C
passed
±0.3 °C
passed
±0. 4 °C
passed
±0.2 °C
passed
±0.3 °C
passed
± 1 . 0 °C
passed
±0.4 °C
passed
±0.4 °C
passed
±0.8 °C
passed
Page 1/1