The pH/mV/Ion electrodes should be connected to the coaxial pH/mV connectors
CH.1, etc. If separate electrodes are used, connect the reference electrodes to the
REF. terminals. Connect the conductivity cell to the coaxial EC input. Automatic
temperature compensation and temperature measurements are possible by plug-
ging a Pt1000 thermocompensator into the °C terminals. Without compensator, the
manual temperature compensation is automatically switched on. C862 & C863: the
O
-electrode should only be connected to the coaxial connector CH.2. Its banana
2
plugs should be inserted in the °C terminals.
A recorder can be connected to the red (+) and black (-) terminals. Use only labora-
tory recorders with a high input impedance!
A standard RS232 output terminal (DP9) is provided for interfacing the instru-
ment with a printer or computer. Data is sent in the ASCII code at a BAUD rate of
300...19200 bps (8 bit, no parity, 1 stopbit).
Serial port pinout specifications:
pin 1 : connected to pin 4 and pin 6
pin 2 : TxD, transmit data
pin 3 : RxD, receive data
pin 4 : connected to pin 1 and pin 6
pin 5 : Gnd, signal ground
pin 6 : connected to pin 1 and pin 4
pin 7 : connected to pin 8
pin 8 : connected to pin 7
pin 9 : not connected
A USB port is provided for interfacing the instrument with a computer. Download
the necessary driver from www.consort.be (>downloads >software).
1. Select [MENU] by pressing MODE.
2. Select [SYSTEM] and press CAL.
3. Follow the instructions on the screen to adjust language and contrast.
1. Select [MENU] by pressing MODE.
2. Select [DATE/TIME] and press CAL.
3. Adjust the date and press CAL to adjust the date.
4. Adjust the time and press CAL to adjust the time.
1. Select [MENU] by pressing MODE.
2. Select [RECORDER] and press CAL.
3. Select the desired channel and press CAL.
4. Select the desired range and press CAL (only for conductivity).
5. Select the desired minimum level corresponding to a 0 V recorder output and
press CAL.
6. Select the desired maximum level corresponding to a 2 V recorder output and
press CAL.
1. Select [MENU] by pressing MODE.
2. Select [PASSWORD] and press CAL.
3. A private code can be programmed to avoid undesired access to the instru-
ment. Choose [YES] and press CAL to enter your secret sequence of 5 keys.
C861 • C862 • C863 • C864 • C868

Inputs

Recorder output

Digital output

Computer interface

System

Date and Time

Recorder

Password