The ProfiSafe 1 is a two-pole voltage tester in
If required, please contact:
accordance with EN/IEC 61243-3 (VDE 0682 part
401) with LED display.
GOSSEN METRAWATT GMBH
You can use the ProfiSafe 1 to measure d.c. and
Service-Center
a.c. voltages between 12 and 690 V.
Thomas-Mann-Strasse 20
Furthermore you can determine polarity, phase and
90471 Nürnberg · Germany
phase sequence and perform continuity tests up
Phone +49-(0)-911-8602-410/256
to 3 MΩ .
Fax
+49-(0)-911-8602-253
The energy is supplied by a high-performance so-
E-Mail service@ gossenmetrawatt.com
lar cell. The integrated lithium battery is permanently
rechargable, even at a minimum light level. This
This address is only valid in Germany. Please
device does not require replacement batteries.
contact our represen≠tatives or subsidiaries for
Thanks to its high degree of protection (IP 65) the
service in other countries.
ProfiSafe 1 may even be used in rain.
Product Support
2.
Safety Precautions
If required, please contact:
You have chosen an instrument which provides you
with a high level of safety.
GOSSEN METRAWATT GMBH
The ProfiSafe 1 voltage tester has been approved
Product Support Hotline
by VDE test authorities for application of VDE GS
Phone +49-(0)-911-8602-112
marking.
Fax
+49-(0)-911-8602-709
When used for its intended purpose, the safety of
E-Mail support@gossenmetrawatt.com
the operator, as well as that of the instrument, is
assured.
In order to maintain flawless technical safety
conditions, and to assure safe use, it is impera-
tive that you read these operating instructions
thoroughly and carefully before placing your
instruments into service, and that you follow all
instructions contained therein.
Observe the following safety precautions:
The voltages indicated on the ProfiSafe 1 are
rated voltages. The voltage tester may only
be used in systems working with those rated
voltage ranges.
Faultless indication of display values is only
guaranteed between -10˚ and +55˚C.
Hold the instrument by the handles only, to
avoid covering the display and not touching
the test electrodes before and during tests.
Just before they are used, voltage testers need
to be checked to ensure they function
correctly. Carry out the function test and check
the instrument at a known voltage source -
e.g. a 230 V socket. If the display of one or
several systems fails in the course of checking,
the instrument must not be used again.
The ProfiSafe 1 has a high internal resistance.
In exceptional cases the device reads values
of inductive or capacitive voltages.
The ProfiSafe 1 must not be switched on for
more than 30 seconds.
For the determination of phase conductors
and phase sequence by means of the
accessible electrode, the perceptibility of the
above lights up according to the connected voltage.
display may be impaired, e.g. when using
Polarity is determined as follows: The „+" 12 V-
protection means against direct contact, in
electrode lights up when the test probe marked „+"
unfavourable locations, for example on
is connected to the positive pole.
wooden ladders or insulating floor coverings,
as well as in unfavourable lighting conditions
4.3 Tests with the Accessible Electrode
and in an improperly, grounded a.c. voltage
The ProfiSafe 1 is equipped with a accessible
system.
electrode „Pol-L1-Sensor", and a triangular LED for
The voltage tester may only be stripped by
phase testing and phase sequence indication.
those authorised personnel.
Attention!
Voltage testers must be clean and dry.
While working with the accessible electrode the
perceptibility of the display may be impaired (see
3. Start-Up
chapter 2).
Your instrument is supplied with a solar cell and an
Note!
integrated lithium accumulator. Thus, the ProfiSafe
1 is ready for operation at any time and does not
Tests performed with the accessible electrode are
require replacement batteries.
only possible at systems working with properly
grounded a.c. voltage systems at voltages of at
3.1 Function test
least 165 V to earth.
To test the functions and your own energy source,
Phase Test
put the test electrodes together. The rectangular
The phase conductor is identified by establishing
green LED (continuity test) lights up permanently
contact with either one of the two test probes, and
for approximately 5 seconds. After 5 seconds all
by touching the accessible electrode „Pol-L1-Sen-
9 LEDs flash up five times. Then the device
sor" at the same time. If the triangle lights up the
switches off automatically.
conductor is energized.
Attention!
Phase sequence test
If one of the LED does not light up, don't use the
To determine the phase sequence between two
device any longer.
phases in the 230/400 V 3-phase network by
applying both test electrodes and touching the
Note!
accessible electrode, proceed as follows:
If the green LCD, after putting the test electrodes
Search for the phase conductors using one pole
together, flickers or doesn't light up, you have to
(see phase test).
recharge the storage battery (see chapter 6).
Apply both test electrodes to two phase conductors
(display: approximately 400 V).
4. Testing
Touch the accessible electrode with your finger.
4.1 General Instructions
When the L1 phase is applied to the test probe
Voltage readings have always priority. If there is no
marked + L1 (with display unit) and L2 phase to
voltage (>3.6 V), then the ProfiSafe 1 is in continuity
the other test probe (without display unit), the
test mode.
triangle lights up if rotation is clockwise. If the
triangle does not light up, the direction of rotation
4.2 Testing Voltage and Polarity
is counter-clockwise. If 230 V is displayed instead
of 400 V, the neutral conductor may have been
Attention!
contacted with one of the test probes.
The maximum allowable on-time for voltage testing
is 30 seconds.
4.4 Continuity Test, Diode Test
Alternating Voltage
Place the test electrodes on the lead to be tested.
Both lower 12 V LEDs (+) light up alternatively and
The rectangular LED (green) lights up with a
the LED chain located above lights up according
resistance of 0 to about 200 kΩ. With resistances
to the connected voltage when both test electrodes
of more than 200 kΩ up to approx. 3 MΩ the LED
are placed on a.c. voltage within the nominal
flashes.
voltage range.
The green rectangle lights up when the test probe
Direct Voltage and Polarity
of the ProfiSafe 1 marked „+" is placed on the
When both test electrodes are placed on d.c.
anode of a semiconductor. Otherwise the
voltage within nominal voltage range, one of the
semiconductor is connected in the reverse
lower 12 V LEDs (+) and the LED chain located
direction.
Note!
Continuity tests are automatically cancelled after
5 seconds in order to spare the lithium battery.
Afterwards the device is immediately ready for
operation. Voltage readings always take priority (see
chapter 4.1).
5 Characteristic Values
Bedienungsanleitung
Operating Instructions
Display:
9 LED's for voltage,
continuity, phase and
phase sequence
Rated voltage range:
12...690 V AC/DC in
ProfiSafe 1
steps 12, 50, 120, 230,
400 and 690 V
Frequency range:
0...2000 Hz
Spannungsprüfer
Current (peak value):
3.1 mA (690 V)
Voltage Tester
Input resistance:
223 kΩ
On-time:
30 s
Continuity test:
0...200 kΩ constantly
gleaming,
0.2 ...3 MΩ flashing
Over-voltage category:
CAT IV
Housing:
ABS, IP 65
Connecting cable:
Rubber insulated
flexible cable
H07RN-F, 1m
6 Maintenance
6.1 Testing and charging the storage battery
To test the functions and your own energy source,
hold the test electrodes together.
The rectangular green LED (continuity test) must
light up for 5 seconds.
If the green LED does not flash or light up when
putting together the test probes, you have to rech-
arge the battery. To recharge the battery, place the
device in a light source (e. g. on a window sill) with
the display field facing up. Due to the high-
performance solar cell, the ProfiSafe 1 will be fully
functional within a short period of time ( approx. 5
minutes).
Perform the function test before putting the
instrument back into operation (see 3.1).
Attention!
Please do not use halogen lamps for charging,
otherwise the device could be damaged.
6.2 Housing
The voltage tester should be kept dry and clean.
The plastic housing can be cleaned with a cloth
GOSSEN METRAWATT GMBH
dampened with isopropyl alcohol or soapy water.
Thomas-Mann-Straße 16-20
D-90471 Nürnberg
Telefon +49-(0)-911 8602-0
Telefax +49-(0)-911-8602-669
www.gossenmetrawatt.com
3-348-971-15
3/1.04

Werbung

loading

Inhaltszusammenfassung für Gossen MetraWatt ProfiSafe 1

  • Seite 1 Voltage readings always take priority (see GOSSEN METRAWATT GMBH You can use the ProfiSafe 1 to measure d.c. and as well as in unfavourable lighting conditions 4.3 Tests with the Accessible Electrode chapter 4.1).
  • Seite 2 4.3. Prüfen mit der Berührungselektrode Spannung, Durchgang, Lithium-Akkumulators zerlegen. Ausland stehen unsere jeweiligen Vertretungen Der ProfiSafe 1 ist durch seine hohe Schutzart (IP Der ProfiSafe 1 ist mit Berührungselektrode „Pol- Phase und Spannungsprüfer müssen trocken und sauber oder Niederlassungen zur Verfügung.