Herunterladen Inhalt Inhalt Diese Seite drucken

VEGA SOLITRAC 31 Betriebsanleitung Seite 71

Vorschau ausblenden Andere Handbücher für SOLITRAC 31:
Inhaltsverzeichnis

Werbung

FF descriptor
POINT_LEVEL_AD-
JUST_MODE
RELAY_VALUE_SEL
DIGITAL_IN
DIGITAL_IN_BOOL
ANALOG_IN
FF_CHANNEL_AVAILA-
BLE
FF_CHANNEL_USED
Mapping of Process Value Status
Hex
Quality
0x00
Bad
0x01
Bad
0x0C
Bad
0x10
Bad
0x1C
Bad
0x20
Bad
0x44
Uncertain
0x48
Uncertain
0x4C
Uncertain
0x80
Good (NC)
Mapping of User Error Codes to FF Field Diagnostics
Priority
Description
31
Hardware failure
30
Memory failure
SOLITRAC 31 • Foundation Fieldbus
Rel. In-
Description
dex
54
Point level adjustment mode
55
Relay basic value
56
Frequency of digital input
57
State of digital input
58
Current on analog input
59
Available channels
60
Used channels
Sub-Status
Non-specific
Configuration Error
Device Failure
Sensor Failure
Out of Service
Transducer in MAN
Last Usable Value
Substitute
Initial Value
Non-specific
Condition
Unexpected error
It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,
117, 120
Five or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,
86, 121, 122, 141
On user error codes: 40, 41, 53, 68, 123, 124, 125
Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application
Transducer block is in mode "Manual"
Three or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 29
After startup of device or channel assignment in
AIFB was changed. The channel unit may be unk-
nown until next "Process Data Update"
On user error codes: 13
No errors concerning to channel handling
On user error codes: 0, 33, 45, 71, 126, 127
User Error Codes
40, 41
34, 35, 36, 37
11 Anhang
Unit
-
-
-
-
-
-
-
NE-107 Status
FAILURE
FAILURE
71

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis