Herunterladen Inhalt Inhalt Diese Seite drucken

Kröber O2-4.0 Gebrauchsanweisung Seite 94

Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Kröber O2-4.0
Appendix
Immunity test
radiated HF
interference acc. to
IEC 61000-4-3
94
IEC 60601 test level
3 V/m
80 MHz to 2.5 GHz
Compatibility level
3 V/m
Electromagnetic
environment - regulations
Portable and mobile radio
equipment should not be
used closer to the Kröber
O2 4.0 and the leads, than
the safety distance
calculated by using the
equation applicable for the
transmission frequency.
Recommended safety
distance:
d= 1.2 √P for 80 MHz to
800 MHz
d= 2.3 √P for 800 MHz to
2.5 GHz

Quicklinks ausblenden:

Inhaltsverzeichnis
loading

Inhaltsverzeichnis