Herunterladen Inhalt Inhalt Diese Seite drucken

Kröber O2-4.0 Gebrauchsanweisung Seite 88

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Kröber O2-4.0
Maintenance
Interval
with changing patients
after infections
after service work
after service work
after service work
after service work
Interval
annually
As average life times, we expect:
Article
Kröber O2-4.0
Humidifer
Oxygen hoses, nasal
cannula
Oarse dust filter
Device inlet filter
88
Cleaning work
bracket must be cleaned and disinfected.
Replace the nasal cannula.
WARNING! Health risk!
In order to avoid cross infection, each user of
the Kröber O2 4.0 should wear his/her own
nasal cannula.
After an infection a new nasal cannula should be used
to prevent re-infection.
The Kröber O2 4.0 must be cleaned with a damp cloth
and disinfected after.
Nasal cannula should be disposed of and replaced by a
new one.
Replace the coarse dust filter
Replace the air intake filter.
Inspection
Safety inspection
Life time
5 years min.
1 year
1 month
1 month
1 month