Herunterladen Inhalt Inhalt Diese Seite drucken

Fuse Replacement - Unitest Hexagon 310 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fuse replacement

8.4 Fuse replacementl
Prior to fuse replacement,
ensure that multimeter is dis-
connected from external vol-
tage supply and the other
connected instruments (such
as UUT, control instruments,
etc.).
Exclusively use fuses of volta-
ge and current values in com-
pliance with the technical data
section.
Using auxiliary fuses, in particular
short-circuiting fuse holders is
prohibited and can cause instru-
ment destruction or serious bodily
injury of operator.
70
c Switch of the instrument.
c Loosen the screws on the
instrument rear.
c Lift the case cover.
c Remove the defect fuse.
c Insert new fuse.
c Replace the case cover and
retighten the screws
FS1: 1A / 240V, IR 6kA (Bussmann
TDC180-1A or other with compari-
son value) UNITEST Order No.:
ES1000MAF001
FS2: 13A / 240V, IR 6kA (Bussmann
TDC180-13A or other with compa-
rison value) UNITEST Order No.:
ES13000MAF

Quicklinks ausblenden:

Werbung

Kapitel