OUTP:CONF (?) <CPD> { 1 | 12 | 123 }
This command defines how many outputs is used in basic power and energy AC modes (Pac and
Eac).
 1 – active are channels U1, I1 only
 12 – active are channels U1, U2, I1, I2
 123 – active are all channels
If query is sent, calibrator returns active channels.
Example:
OUTP:CONF 123 <cr> - activates all channels
OUTP: CONF? <cr> - the calibrator returns 123
OUTP:EQU:FACT(?) <DNPD>
This command sets number of pulses per selected unit (pulses generated on Energy pulse output
on the rear panel BNC).
<DNPD>
Parameter represents the value number of pulses per selected unit.
If query is sent, calibrator returns set value using standard exponential format. Example: 400.0
i/kWh is returned as 4.000000e+002.
Example:
Number of pulses 1000 i/kVArh:
OUT:REF:CONS 1000.0
OUT:REF:UNIT VAR
MEASure und CONFigure Subsystem
Bedienung des internen Multimeters.
Keyword
MEASure?
CONFigure(?)
MEAS?
Nach diesem Befehl wird der Messwert vom Multimeter geschickt.
Die Antwort ist im exponentiellen Format.
Beispiel:
5.456 V wird als 5.456000e+000 gesendet
Beispiel:
MEAS? Kalibrator antwortet den Messwert
Owner's Manual
Parameter
<CPD> { VOLTage | CURRent | FREQuency }
OC
M-133C Power Calibrator
70