9.2

Air Temperature Readings

Error Code
Problem
E02
The air temperature value is
under the lower limit.
E03
The air temperature value
exceeds the upper limit.
E31
The temperature sensor or
measuring circuit is dam-
aged.
9.3

Air Humidity Reading

Error Code
Problem
E04
The air temperature meas-
urement has an error code.
E11
The RH calibration has
failed.
E34
The RH sensor or the
measuring circuit is dam-
aged.
Solution
Put the meter in regular room temperature
for 30 minutes, if the error message still
appears, send the device for repair.
Put the meter in regular room temperature
for 30 minutes, if the error message still
appears, send the device for repair.
Send the device for repair.
Solution
Refer to the above mentioned temperature
error code.
Send the meter back for recalibration.
Send the meter back for repair.