10.1 Trouble-shooting (Continuation)
Problem
Insufficient aspiration power.
VacuSafe doesn't function.
Level monitor doesn't function.
Pump doesn't start. No vacuum.
Display runs normal.
28
Cause
The filter is clogged.
Blown fuse or wrong voltage set.
Level monitor is deactivated.
Pump motor overheated because
of continuous running and high
ambient temperature.
Remedy
Replace the filter.
Change the fuse, check the voltage
selection switch.
Activate level monitor.
Avoid unnecessary loss of vacuum
and continuous running of the
pump by using the hand operator
VACUBOY or use a tube clamp.