NB: Always keep the lid tight – otherwise a
pressure cannot be generated in the bottle!
Connect lid (2) and pump (PRESSURE
connection) with tube (4) on the pump side (blue,
with filter)
Ensure that:
- The blue tube connector is fitted to the blue
socket
- The rapid-action connector is properly fitted
(it must click).
NB: Always use the integrated sterile filter on the
pump side (blue). This prevents medium from enter-
ing the pump and causing damage.
Connect the VACUBOY hand operator tube or
other tube (3) to the tube connector on the red
rapid-action connector.
3. Dispensing
Check that the venting screw is closed.
Switch on the pump with the power switch
Set values for pressure and hysteresis if necessary
The pump will begin to function.