NB: Always keep the lid tight – otherwise pressure
cannot be generated in the bottle!
Connect lid (2) and pump (PRESSURE
connection) with tube (4) on the pump side (blue,
with filter)
Ensure that:
- The blue tube connector is fitted to the blue
socket
- The rapid-action connector is properly fitted
(it must click).
NB: Always use the integrated sterile filter on the
pump side (blue). This prevents medium from enter-
ing the pump and causing damage.
Connect the VACUBOY hand operator tube or
other tube (3) to the tube connector on the red
rapid-action connector.
2. Prepare the vacuum bottle
Ensure that no siphon tubing is connected to the
lid.
NB: When aspiration is carried out using a siphon
tube, air bubbles will rise through the medium
during the aspiration process.
Place lid (5) on the vacuum bottle and ensure that
it fits tightly.
NB: Always keep the lid tight – otherwise a vacuum
cannot be generated in the bottle!
Connect lid (5) and pump (VACUUM connection)
with tube (8) on the pump side (blue, with filter)
Ensure that:
- The blue tube connector is fitted to the blue
socket
- The rapid-action connector is properly fitted
(it must click)
19