INTEGRA Biosciences VacuSafe Bedienungsanleitung Seite 20

Absaug-system
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

NB: Always use the integrated sterile filter on the
pump side (blue). This prevents medium from enter-
ing the pump and causing damage.
Connect the VACUBOY hand operator tube or
other tube (6) to the tube connector on the red
rapid-action connector.
Connect sensor cable (7) between lid (5) and pump
(connection SENSOR).
3. Aspirating and dispensing
Procedure:
Connect foot switch (9) to the FOOTSWITCH
socket of the pump
Ensure that the venting screw on both bottles are
closed
Switch on the pump with the power switch
If necessary, set values for pressure and vacuum
The pump will begin to function
20
Either the foot switch or the MODE key can be
used to switch between "aspirate" and "dispense"
Once the bottle is full, the level indicator will be
activated – the red LED SENSOR will blink. The
bottle must be emptied before proceeding further.
NB: For proper operation of the level indicator, and
to prevent the build-up of foam, we recommend the
use of normal commercial anti-foaming agents (for
example Antifoam A from Fluka or Sigma)