NB: When aspiration is carried out using a siphon
tube, air bubbles will rise through the medium
during the aspiration process.
Place lid (1) on the vacuum bottle and ensure that
it fits tightly.
NB: Always keep the lid tight – otherwise a vacuum
cannot be generated in the bottle!
Connect lid (1) and pump (VACUUM connection)
with tube (4) on the pump side (blue, with filter).
Ensure that:
- The blue tube connector is fitted to the blue
socket
- The rapid-action connector is properly fitted
(it must click).
NB: Always use the integrated sterile filter on the
pump side (blue). This prevents medium from enter-
ing the pump and causing damage.
Connect the VACUBOY hand operator tube or
other tube (2) to the tube connector on the red
rapid-action connector
Connect sensor cable (3) between lid (1) and pump
(SENSOR).
3. Aspiration
Check that the venting screw is closed
Switch on the pump with the power switch
Set values for vacuum and hysteresis if necessary
The pump will begin to function
Once the bottle is full, the level indicator will be
activated – the LED SENSOR will blink and an
acoustic alarm starts. The pump stops. The bottle
must be emptied before proceeding further.
NB: For proper operation of the level indicator, and
to prevent the build-up of foam, we recommend the
use of normal commercial anti-foaming agents (for
example Antifoam A from Fluka or Sigma)
15