Problem
Causes and solutions
The error code "E7"
The app was not successfully installed on your
and the Bluetooth®
smartphone or tablet.
indicator appear on
•
Reinstall the app and log in with your user
the display.
data.
Your smartphone or tablet does not support
Bluetooth® Smart.
•
Exchange your smartphone or tablet with
a model that supports Bluetooth® Smart.
Not all
You performed more than 60 measurements
measurements saved
without connecting to the app.
are transmitted.
The blood pressure monitor can save a total of
60 measurements per user. Afterwards, the
first measurement will be overwritten by the
last one.
•
Connect the blood pressure monitor with
the app on a regular basis to prevent
losing measurements.
Use the FAQs (frequently asked questions) on our website as an
additional help for your scale: http://www.krippl-watches.com
44
Technical data
Type:
Model:
EAN:
Power supply:
Classification
Type of operation
Protection class
IP code
Anaesthesia safety
Measurement method
Measurement accuracy
Measurement range
Cuff pressure:
Systolic pressure:
Diastolic pressure:
MD4000
B2B-BPM-1
9020400119677
Batteries, 4 x 1.5 V DC,
type LR03/AAA
Type BF application part
Continuous operation
Internal power supply
IP21
The device is not approved in
accordance with category AP or APG.
Oscillometric, non-invasive
measurement of blood pressure on
the upper arm
Maximum permissible standard
deviation according to clinical
examination: systolic 8
mmHg/diastolic 8 mmHg
30 – 250 mmHg
50 – 250 mmHg
30 – 200 mmHg
45