Problem
Causes and solutions
The error code "E5"
A problem has occurred when inflating the cuff
appears on the
(inflation pressure > 300 mmHg).
display.
•
Make sure that the cuff hose is not kinked
or clamped.
•
Fit the cuff correctly on your arm.
•
Repeat the measurement.
The display is blank,
The batteries were not inserted correctly.
or the battery
•
Remove the batteries and put them back
indicator and the
in with the correct polarity.
error code "E6"
Batteries are empty.
appear on the
•
display.
Replace the empty batteries with new
ones.
42
Problem
Causes and solutions
The error code "E7"
Bluetooth® transfer is not possible.
and the Bluetooth®
Once you have prepared everything for
indicator appear on
automatic transmission, repeat the
the display.
measurement or transfer the measurements
manually.
Bluetooth® transfer is not possible.
The Bluetooth® function of your smartphone
or tablet is switched off.
•
Switch the Bluetooth® function of your
smartphone or tablet on.
The Bluetooth® transmitter of the blood
pressure monitor is not working properly.
•
Take the batteries out of the battery
compartment.
•
Put the batteries back in.
•
Switch the Bluetooth® function of your
smartphone or tablet off and back on.
Your smartphone or tablet is not working
properly.
•
Switch off your smartphone or tablet off
completely and then back on.
43