Malfunction
Er:X:16
...
is displayed?
... when the Enter key
(send/print) is pressed
Er:X:71
and
played?
... when the Enter key
(send/print) is pressed
Er:X:72
and
played?
10.MAINTENANCE/RECALIBRATION
10.1 Information about maintenance and recalibration
10.2 Check content of calibration counter
Oscillation of the scales has occurred, preventing
determination of the zero point.
- Restart scales
Data transmission not possible, wireless module is
disabled.
is dis-
- Activate wireless module (see "Activate wire-
less module (SYS)" on page 57).
Data transmission not possible, no wireless group
set up.
- Set up wireless group (see "Set up wireless
is dis-
group (Lrn)" on page 58).
We recommend having maintenance performed prior
to recalibration of the scales.
Ensure that recalibration is carried out by authorised
persons in accordance with the national legal require-
ments.
Recalibration is required in all cases if one or more cali-
bration seals have been damaged or the content of the
calibration counter does not match the number on the
valid calibration counter sticker.
These seca scales have been calibrated. Calibration
may only be performed by authorised bodies. To
ensure this, the scales are equipped with a calibration
counter which registers every change in data relevant
to calibration.
Cause/Remedy
ATTENTION!
Incorrect measurement due to improper
maintenance
– Make sure that maintenance and repair are
only carried out by an authorised service
partner.
– You can find your local service partner at
www.seca.com or just send an e-mail to
service@seca.com.
Maintenance/Recalibration • 63
English