Water level alarm
Inhaltsverzeichnis

Werbung

Failure: One of the red LEDs (1) is lit; the alert
signal is sounded; the switched socket outlet
has shut down; however, the sensors are in a
position which should not trigger an alarm.
Cause: The float ring of the sensor was turned by
180° during the last cleaning process. Thus, the
switching direction has changed.
Remedy: Turn the float ring of the sensor into the
correct position. Please also refer to the chapter
titled "The sensors".
Failure: Frequent switching operation of the
Water Level Alarm
reason.
Cause: Another electrical unit is interfering with the
function of 7607/2.
Remedy: Remove or switch off the interfering unit.
Cause: The 7607/2 is defective.
Remedy: Remove the unit or send to the dealer or
manufacturer for repair.
without recognisable
Disfonctionnement: Une des LED rouges (1)
s'allume, le signal sonore retentit, la prise
commandée est mise hors service. Cependant,
les capteurs ne se trouvent pas dans une
position entraînant une alarme.
Raison: Lors du dernier nettoyage, un flotteur de
capteur s'est trouvé inversé de 180°. Le sens de
commutation est inversé.
Solution: Repositionnez correctement le flotteur du
capteur en question, voir aussi «Capteurs...»
Disfonctionnement: Commutations fréquentes
de Water Level Alarm sans cause apparente.
Raison:
Un
appareil
fonctionnement de 7607/2.
Solution: Eloignez ou débranchez l'appareil en
cause.
Raison: 7607/2 est défectueux.
Solution: Remplacez ou faites réparer l'appareil
auprès de votre revendeur ou fabricant.