Specific warning
• Avoid repeated intracerebral catheter im-
plantations: Frequent brain perforations to
permit catheter insertions may predispose
to intracerebral oedema and haemorrha-
ge which can lead to a rise in intracranial
pressure.
• The intracranial screw is made from a high
performance polymer material, which has
advantages over metal when used in ima-
ging processes. The user must ensure that
the polymer screw is not subjected to any
great lateral force, especially impacts.
• Do not use excessive force when turning
the intracranial screw.
• The BOLT screw has a mechanical stop
that allows it to be screwed in up to 8 mm
(BOLT CH5) or 8.5 mm (CH9). The depth of
insertion for secure fixation and application
of the BOLT screw in the cranium is up to
the physician in charge.
• When using the CH 5 RAUMEDIC
(-DRILL) KIT, only catheters with an outer
diameter of CH 5 should be used.
• When using the CH 9 RAUMEDIC
(-DRILL) KIT, only catheters with an outer
diameter of CH 9 should be used.
• Only NEUROVENT
®
VENT
-TO catheters may be used with the
®
RAUMEDIC
BOLT (-DRILL) KIT PTO.
®
BOLT
®
BOLT
®
PTO and NEURO-
RAUMEDIC
– BOLT KIT / DRILL KIT
®
• Only NEUROVENT
-PTO 2L and µDialysis
®
MD70 BOLT version catheters may be used
with the RAUMEDIC
®
• When using the RAUMEDIC
PTO 2L with a microdialysis catheter, the
instructions for use of the MD system must
also be observed.
• Only the catheters NEUROVENT
and NEUROVENT
with sleeve housing
®
may be used with RAUMEDIC
DRILL KIT VP 16 and the fixing cap must
always be closed up to the stop hereby.
• After application, check that the measure-
ment system used with the RAUMEDIC
BOLT KIT is functioning correctly.
• Measurements should only be taken if the
screw is firmly in place.
• Warning: once a system is implanted, a re-
placement system should always be availa-
ble should the need arise.
• Due to the increased risk of infection with
longer applications, a change after 5 days
and no later than 10 days is recommended.
It is up to the user to assess the risk of in-
fection if using the RAUMEDIC
for longer periods.
BOLT KIT PTO 2L.
BOLT KIT
®
VP 16
®
BOLT-
®
®
BOLT KIT
®
11