Before measurements can be recorded properly, it is necessary to
define the recorder-span (values corresponding to the left and right
corners of the graphic paper). The displayed possibilities depend from
the channel set-up.
1.
Press SET.
2.
Press
or
continue or SET to cancel.
3.
The display shows e.g. [rc1=EC1], [rc1=EC2], [rc1=°C1], or
[rc1=°C2]. Select the desired channel & mode for recorder-1 with
or
and press CAL to continue or SET to cancel.
4.
The display shows e.g. [rc=3.00_] while = is blinking. Choose the
value corresponding to a 4 mA recorder-1 output with
press CAL to continue or SET to cancel.
The display shows e.g. [rc=10.00 _ ] while = is blinking. Choose the
5.
value corresponding to a 20 mA recorder-1 output with
press CAL to continue or SET to cancel.
6.
The display shows e.g. [rc2=EC1], [rc2=EC2], [rc2=°C1], or
[rc2=°C2]. Select the desired channel & mode for recorder-2 with
or
and press CAL to continue or SET to cancel.
7.
The display shows e.g. [rc=7.00_] while = is blinking. Choose the
value corresponding to a 4 mA recorder-2 output with
press CAL to continue or SET to cancel.
The display shows e.g. [rc=13.00 _ ] while = is blinking. Choose the
8.
value corresponding to a 20 mA recorder-2 output with
press CAL to continue or SET to cancel.
The measuring range is common for both channels and can be
programmed out of 6 possibilities: 20 (0.1 cm
2 - 20 - 200 - 2000 (10 cm
1.
Press SET.
2.
Press
or
continue or press SET to cancel.
3.
The display shows [rEF 25°C] or [rEF 20°C]. Select the temperature
to which all future measurements will be referred with
press CAL to continue or press SET to cancel.
4.
The display shows [cc=0.1], [cc=1.0] or [cc=10]. Select the cell
constant of the connected cell with
continue or press SET to cancel.
5.
The display shows e.g. [r=200.0µS], [r=2000µS], [r=20.00MS], or
[r=200.0MS] for a 1 cm-1 cell. Select the desired range for both
channels with
cancel.
until the display shows [SEt rEc] and press CAL to
cell only) mS/cm.
-1
until the display shows [Set EC] and press CAL to
or
and press CAL to continue or press SET to
or
or
or
or
cell only) - 200 µS/cm -
-1
or
or
and press CAL to
Recorder
and
and
and
and
Range select
and
5