Interfacing with a printer:
Run the RS232 set-up procedure for automatic printing at timed intervals
during the process control. The result looks e.g. as follows:
#005
705 µS1/cm , 21.4 °C (/H) (/P)
853 µS2/cm , 54.3 °C (/L) (/O)
#006
361 µS1/cm , 30.9 °C (/H) (/W)
578 µS2/cm , 15.0 °C (/L) (/A)

TROUBLE SHOOTING

This procedure resets the instrument to the original factory settings.
1.
Switch the instrument ON while pressing SET.
2.
The display shows [Ec=----] while [=] is blinking. Enter the engineer-
code (SET, CAL,
the original manufacturer settings.
3.
The display shows [rESEt] for a while and returns to the
measurements.
(H=high-alarm, P=proportional)
(L=low-alarm, O=on/off)
(H=high-alarm, W=wait-phase)
(L=low-alarm, A=general alarm)
, SET) to reset all parameters in the memory to
Printer
connection

Reset

13