A standard RS232 connection is provided for interfacing the instrument
with a computer, printer or other R300's. Data is sent or received in the
ASCII code at a programmable BAUD rate of 150...4800 b/s (8 bit, no
parity, 1 start and 2 stopbits, automatic line feed ON, data transfer protocol
X-ON/X-OFF).
TxD, transmit data to terminal (+) of (IN)(RS232).
TxD, signal ground to terminal (-) of (IN)(RS232).
RxD, receive data to terminal (+) of (OUT)(RS232).
RxD, signal ground to terminal (-) of (OUT)(RS232).
Interfacing with a computer:
Any desired command can be included in a computer data acquisition
program. Follow exactly the same measuring or calibration procedures as
described in this manual. The instrument is able to recognise commands
in two ways:
Simple: the simple method can be used when only one controller is
connected to the computer (identification number should be zero!). Send
a single character to the instrument and it will immediately execute the
corresponding command as follows:
1 = press SET.
2 = press
.
3 = press
.
4 = press CAL.
? = send display to computer.
+ = keyboard on (manual operation remains possible).
-
= keyboard off (manual operation disabled).
V = start the input of a value (only valid in certain routines).
This example makes the computer press SET in the simple mode:
10 OPEN "COM1:2400,N,8,2" AS #1 'prepares communication port
20 PRINT #1,"1"
Sophisticated: this method is used for all types of communications such
as a network of several R300's connected to the computer. The
commands should always be preceded by the identification number of the
controller concerned (see Set-up):
#xxx = "#" + identification number (xxx) in ASCII code. All blank (ASCII-
32) or CR (ASCII-13) characters following the identification
number are ignored!
C1
= press SET.
C2
= press
.
C3
= press
.
C4
= press CAL.
C?
= send display to computer.
C+
= keyboard on (manual operation remains possible).
C-
= keyboard off (manual operation disabled).
V
= start the input of a value (only valid in certain routines).
LF
= linefeed (ASCII-10), closes a command or input and permits the
instrument to receive a next command.
'SET key pressed

Digital output

Computer
connection
11