1.
Select the desired °C channel with
2.
Measure the real °C value of processed liquid with an other electrode
and meter.
3.
Press
or
until the display shows [ShiFt] and press CAL to start
the automatic calibration procedure.
4.
The display shows e.g. [°C=22.3]. Calibrate to the proper value with
or
and press CAL to return to the measurements. Press SET
here if you wish to return to the original readings. The display shows
[rESEt?]. Press CAL to reset the shifted values or press SET to
cancel.
∗ When the selected channel has been programmed to control °C, it
cannot be calibrated unless a Pt1000 (or Pt100) is connected to the
proper input.
∗ Without Pt1000 (or Pt100), change the manual temperature
compensation with
measurements.
DATA COMMUNICATION
This procedure permits to recall some important values recorded during
the measuring & calibration phases of the controller. These values are
reset after each new calibration!
1.
Press SET.
2.
Press
or
continue or SET to cancel.
3.
The display shows the cell constant of the conductivity cell connected
to channel-1, e.g. [cc1=1.056]. Press CAL to continue or SET to
cancel.
4.
The display shows the cell constant of the conductivity cell connected
to channel-2, e.g. [cc1=0.981]. Press CAL to continue or SET to
cancel.
5.
The display shows the lowest value ever measured by channel-1, e.g.
[Mi=635]. Press CAL to continue or SET to cancel.
6.
The display shows the highest value ever measured by channel-1,
e.g. [MA=807]. Press CAL to continue or SET to cancel.
7.
The display shows the lowest value ever measured by channel-2, e.g.
[Mi=312]. Press CAL to continue or SET to cancel.
8.
The display shows the highest value ever measured by channel-2,
e.g. [MA=381]. Press CAL to continue or SET to cancel.
Two recorders, one for each channel, can be connected to the (+) and (-)
terminals (OUT1) and (OUT2). The 4 to 20 mA output range corresponds
with the pre-set values (see 'Recorder' page 5).
or
and press CAL to return to the
until the display shows [diAGn] and press CAL to
or
and press CAL.
Rapid °C shift
Diagnosis
Analogue
output
10