Template: Bridge Sensor
Parameter
Calibration Period
(Days)
Measurement
location ID
1)
Typical values for an HBM T40FH/150 kN@m torque flange
Template: HBM Channel Name
Channel name
When creating the Bridge Sensor template, the manufacturer defines the physi
cal measured quantity and the physical unit.
The available unit for the particular measured quantity is specified in the IEEE
Standard. For the measured quantity of torque, the unit is "N@m".
At the time of creating the template it is also necessary to choose between the
options "Full Precision", "mV/V" and "uV/V" for the accuracy of the characteris
tic transducer curve mapped in TEDS.
The factory setting is "Full Precision", in order to be able to use full digital reso
lution. This choice is also recommended to users who program the TEDS mem
ory themselves.
T40FH
1
)
Value
Unit
730
days
0
A4429-3.0 HBM: public
TEDS transducer identification
Required
Explanation
user
rights
CAL
Time before recalibra
tion, calculated from the
date specified under Cal
ibration Date.
USR
Identification number for
the measuring point.
Can be assigned accord
ing to the application.
Possible values: a num
ber from 0 to 2047. If this
is not enough, the HBM
Channel Comment tem
plate can also be used
for this purpose.
T40FH/150 kNm
43