Intended Use
The Motion Foot
SLX hydraulic prosthetic foot is intended for use in lower
®
extremity prostheses . The Motion Foot SLX sets itself apart by combining a
two-spring prosthetic foot design with a hydraulic mechanism that allows for
plantar and dorsiflexion range of motion. Fillauer has designed the Motion
Foot SLX to support each phase of foot function in the gait cycle—from heel
strike, to mid-stance, to toe off. The result is a foot that adapts well to sloped
conditions, creating early stance phase stability with solid energy return . This
provides unsurpassed smoothness, stability, and performance from initial
contact, to mid-stance, to terminal stance .
Figure 1
Indications
• Moderate to active transtibial or transfemoral amputees as defined by
functional K3 activity levels
• Unilateral or bilateral patients
• Patients that would benefit from increased flexibility and smooth rollover
• Patients weighing up to 275 lbs . (125 kg)
Contraindications
• Clearance below 4 .08 in . (10 .36 cm)
• Patients weighing over 275 lbs . (125 kg)
• Patients wanting to run or jog regularly on the device
The device is intended for single patient use only .
Performance Characteristics
• Patient weight: Up to 275 lbs . (125 kg)
• Foot weight: 23 .8 oz . (675 g)
• Build height: 4 .08 in . (10 .36 cm)
3