Table 3
Guidance and manufacturer's declaration – electromagnetic
immunity
The Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of The Pulse
Oximeter should assure that it is used in such an electromagnetic
environment.
Immunity
test
Conducted
RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
IEC60601
Complian
test level
ce level
3 Vrms
150 kHz to
N/A
80 MHz
3 V/m
3 V/m
80 MHz to
2.5 GHz
Electromagnetic environment
-guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of The Pulse
Oximeter, including cables,
than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
P
d=1.2
P
d=1.2
P
d=2.3
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m). b
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
18
80MHz to 800MHz
800MHz to 2.5GHz