transient/
burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field
IEC61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test
level.
power
Supply lines
±1 kV for
input/output
lines
±1kV line (s)
to line(s)
±2kV line(s) to
earth
<5% UT
(>95% dip in
UT) for 0.5
cycle
<40% UT
(60% dip in
UT) for 5
cycles
<70% UT
(30% dip in
UT) for 25
cycles
<5% UT
(>95% dip in
UT) for 5 s
3A/m
17
N/A
N/A
3A/m
N/A
N/A
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.