TROUBLESHOOTING
PROBLEM
Tip dripping
Faint display; reduced speed.
Instrument not responding.
Instrument not responding.
Instrument not responding.
ERROR MESSAGE DISPLAYED
No message
Battery symbol flashing
ERR 001
ERR 002
ERR TIP
POSSIBLE CAUSE
Leaking tip
Battery low
●
Locking lever not engaged
during aspirating or
.
dispensing
●
Locking lever open at lower
third of the tip.
Motor blocked
●
Battery low.
●
Viscosity is too high
(see page 7).
●
Tip orifice obstructed.
Locking lever opened too early
within the lower third of the tip.
CORRECTIVE ACTION
Replace tip
Recharge battery pack or replace with charged battery
pack.
1.
Press ENTER to confirm error.
Follow MENU instructions to avoid instrument damage!
Display: CLO (Close).
2.
Engage the locking lever.
Display: rEF (Reference run).
3.
Press the STEP key.
Attention: Liquid remaining in tip will be discharged during
the reference run (rEF). Discharge sample into a suitable
vessel (see pages 5, point 10).
1.
Press ENTER to confirm error.
Follow MENU instructions to avoid instrument damage!
●
Recharge or replace battery pack.
●
Reduce aspirating/dispensing speed.
Display: rEF (Reference run).
Press the STEP key
2.
Attention: Liquid remaining in tip will be discharged
during the reference run (rEF). Discharge sample into a
suitable vessel (see pages 5, point 10).
1.
Press ENTER to confirm error.
Follow MENU instructions to avoid instrument damage!
2.
Display: OPE (Open).
Push the locking lever to its upper stop.
|