Final Fitting Checklist
A properly applied device will look like Fig. 11.
• Collar extends from mandible to just below sternal notch.
• Chin is centered comfortably in Chin Support. Chin should not extend
over edge of Sorbatex™ pad, nor fall inside collar.
• Straps are aligned to the Hook landing area and are an equal length.
• Sides of Back should overlap sides of Front.
• Front of collar angled up toward ears. Lower plastic edge should not
be resting on patient´s clavicles nor digging into trapezius.
• No plastic touching skin. Sorbatex™ pads extend beyond all plastic
edges.
• No slack or gaps in any of the straps.
• Tracheal opening and posterior vent are midline.
• Collar sits away from the neck. There should be a "finger" gap between
the tracheal opening and the neck.
If the collar is fit too closely/tightly to the neck, size up to the next taller
size.
Accessories and Replacement Parts
Please refer to the Össur catalog for a list of available replacement parts
or accessories.
USAGE
Cleaning and care
Remove the pads from the plastic shells (Fig. 12)
• Hand-wash using mild detergent and rinse thoroughly.
• Wring out the excess water and squeeze in a towel. Lay flat to air dry.
Note: Do not machine-wash, tumble dry, iron, bleach, or wash with fabric
softener.
Note: Avoid contact with salt water or chlorinated water. In case of
contact, rinse with fresh water and air dry.
Device plastic can be washed with a damp cloth and mild soap.
• Replace the pads by attaching the grey/dull side to the hook on the
inside of the plastic shells.
DISPOSAL
The device and packaging must be disposed of in accordance with
respective local or national environmental regulations.
LIABILITY
Össur does not assume liability for the following:
• Device not maintained as instructed by the instructions for use.
• Device assembled with components from other manufacturers.
• Device used outside of recommended use condition, application, or
environment.
9