8 Troubleshooting Guide

Situation
Power Failure
Unit Not Heating
Heating, but unit not reaching
temperature or temperature is
fluctuated
ERROR 1
ERROR 3
If other problems arise, or one of the above not solves problems, please contact your
official agent or the manufacturer.
Reparations and amendments should only made by
Contraventions exclude the warranty.
witeg Labortechnik GmbH, WAC-R Steam Sterilizer
29
Confirmations and Solutions
 Check the supplied voltage is present.
 Check the power connection and the power cord for
proper connection and integrity.
 Check if the MAIN Power Switch is "ON".
 Fuse is blown.
 Check if the MAIN Power Switch is "ON".
 Check if the set temperature (SV) confirmed with the Jog-
Dial knob (by turn the Jog-Dial knob cannot change the
set temperature – see point 6.4)
 Check the Safety switch is set higher (min. 10°C-20°C)
than the set temperature (SV).
 Check the Safety switch is set higher (min. 10°C-20°C)
than the set temperature (SV).
 Problem in the sensor or main controller.
 Temperature Sensor Error.
 Low level error. Please fill water in.