occurred, and then the Error message is printed. However, if an Error is occurred while
proceeding till the sterilization stage, the Error details is printed out, and if an Error is
occurred in the initial process, it is not printed. ex) Error_#1 : 2014.03.07_12:34:56
6.6.1 Set Date & Time
 To reach this menu you have to use the SUB button in the menu item "set printer".
 Set up the current date and time to record on the printout paper of the printer.
 If indicating the standby mode while the cursor is blinking, the user can change the setting
value by turning the Jog-Dial Knob to the right/left.
 If pressing the PUSH button, the Displayed value is stored, and the cursor will be moved to
the next position.
 If pressing and holding the PUSH button after setting both DATE and TIME, the values are
stored, fixing while blinking the screen of Display.

6.6.2 Set Name

 To reach this menu you have to use the SUB button in the menu item "Set Date & Time".
 The title of experiment or the sterilization information to record on the printout paper of the
printer can be entered.
 If indicating the standby mode while the cursor is blinking, the user can change the setting
character and numbers by turning the Jog-Dial Knob to the right/left.
 If pressing the PUSH button, the Displayed value is stored, and the cursor will be moved to
the next position.
 If pressing and holding the PUSH button after setting the Name, the values are stored,
fixing while blinking the screen of Display.
 If pressing the MAIN button after completing all the settings, the screen is converted to the
Set Printer mode.
 Press the MAIN button to leave the "set name" menu. Than you are in the menu item "set
printer".

6.7 Locking mode

 To avoid accidental changing of the set values, push the MAIN button in the menu item
"set printer".
 In this menu item can you see the actually temperature in the SV-line and the remaining
time of turn-off delay in the PV-line. The function of the Jog-Dial knob is deactivated.
 Press the MAIN button once more to leave the key lock. Than you are in the menu item
"set turn-off delay".

6.8 Set Operation Cycle Mode

WAC-R Steam Sterilizer
LIQUID cycle mode. As the name implies, LIQUID cycle is for autoclaving liquids while the
SOLID cycle is for autoclaving solid or dry materials. The following table shows you the
difference of the two cycles and proper use:
Type of
Appropriate
Cycle
Material
SOLID
Dry goods,
or "fast
glassware,
exhaust"
etc.
LIQUID
of "slow
Liquids
exhaust"
26
offer two different operation cycles named SOLID cycle mode and
This cycle charges the chamber with steam and holds it at a set
pressure and temperature for a set period of time. At the end of the
cycle, a valve opens and the chamber rapidly returns to atmospheric
pressure.
This cycle prevents sterilized liquids from boiling. Steam is
exhausted slowly at the end of the cycle, allowing the liquids (which
will be super-heated) to cool.
Description
witeg Labortechnik GmbH, WAC-R Steam Sterilizer
www.witeg.de