www.witeg.de
The delay time is adjustable from one minute till over 99 hours in one minute steps.
Turn the Jog-Dial knob to set your delay time. Push the Jog-Dial if you reached the desired
value to confirm it. With the confirmation the timeframe starts elapsing.
If you want to change the value after confirmation, you have to push the Jog-Dial once
again. The PV initiates blinking and you can modify the value again.
To confirm the value and start the delay time push the Jog-Dial yet again.
6.4 Set temperature
When you push the MAIN button in the menu item "set turn-off delay" you reach the menu
item "set temperature". You can recognize this menu at "°C" in SV and PV line.
The temperature is adjustable from 25°C till 132°C in 0,1°C-steps.
Push the MAIN button once again if you want to work without the heating function. Then
you reach the next menu without collimated temperature value.
Choose your desired temperature by turning the Jog-Dial and confirm the set value by
pushing the Jog-Dial.
For changing the confirmed value again, you first have to push the Jog-Dial knob again.
"HEATING" starts blinking at the LCD and the temperature is adjustable again.
When the heating lamp is shining, the heater is activated.
6.5 Temperature Offset
If the temperature measured by you differs from the temperature the device is showing (PV) you
have the possibility to compensate the divergence by the Offset. So you can balance
differences of about ±10 °C.
The Offset is calculated as follows:
Temperature measured by you – showed temperature (PV) = Offset
To reach this menu you have to use the SUB button in the menu item "set temperature".
Set the calculated value by turning the Jog-Dial. The Offset will be added to the PV
immediately.
Save the offset by holding the Jog-Dial pushed until the values in PV and SV line starts
blinking.
Push the MAIN button to leave the justification menu.
6.6 Set Printer
When you push the MAIN button in the menu item "set temperature" you reach the menu
item "set printer".
Turn ON/OFF the output functions of the printer that can record the sterilization operation
process.
If pressing the PUSH of Jog Dial Knob, the Display of ON/OFF of the printer function is
changed. "Set to Print" and "Set not to Print" of the Display are alternately changed every
time when the button is pressed.
If pressing and holding the PUSH button, the set screen is stored, fixing while blinking the
words of the Display.
Printing is automatically carried out at 15 minutes (end of exhaust stage) after completing
1 Cycle (80°C basis) of normal sterilization.
If pressing the Door Open button at the exhaust stage, the sterilization operation process is
printed every time you press the button. However, if the user prints using the button, the
automatic printing stage for 15 minutes is omitted.
If an Error is occurred during operating, print the Data till immediately before the Error is
witeg Labortechnik GmbH, WAC-R Steam Sterilizer