INSTALLATION
1.
Tighten the threaded connector firmly and seal with a suitable
sealant.
CAUTION! Tighten the threaded connector by hand, and then
turn an additional quarter turn using a spanner. The connection
must not be forced. If it is screwed tight with too much force,
the threaded connector can be constricted through radial
compression, which locks the probe tip in a fit that is too tight
and can damage it irrevocably.
2. Remove the protective cap from the probe.
CAUTION! The probe tip must not be damaged.
3. Measure the mounting depth as shown in the figure below.
Mark this distance on the probe.
CAUTION! The probe tip must be positioned in the flow.
32
+/- 5 mm