4.2 Limits of reprocessing
Limits of reprocessing in a home environment
Nebuliser and accessories, disinfection
PARI BABY mask
and PARI BABY bend,
Disinfection
Limits of reprocessing in a professional healthcare institution
Nebuliser and accessories, disinfection
Nebuliser and accessories, sterilisation
Connection tubing
PARI BABY mask and PARI BABY bend,
disinfection
PARI BABY mask and PARI BABY bend,
sterilisation
4.3 Preparation for reprocessing
Procedure
• Detach the connection tubing from the compressor.
• Detach the connection tubing from the nebuliser.
• Detach the mouthpiece from the nebuliser.
• Make sure that all residual volume is removed from the nebuliser.
• Dismantle the nebuliser into its individual parts.
• Carefully pull the blue exhalation valve out of the slot in
the mouthpiece. The exhalation valve must still be at-
tached to the mouthpiece.
• Disconnect all mask components from the nebuliser.
• Detach the PARI BABY bend from the mask if applicable.
300 processing cycles, max. 1 year
300 processing cycles, max. 1 year
300 reprocessing cycles, max. 1 year
100 processing cycles, max. 1 year
50 reprocessing cycles,
max. 1 year
300 processing cycles, max. 1 year
100 processing cycles, max. 1 year
29